Radiofrequency coil design for improving human liver fat quantification in a portable single-side magnetic resonance
Shiwei Yang1,2, Xiao Chen3, Suen Chen1,2
1School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.
Abstract:
Earlier diagnosis of nonalcoholic fatty liver disease (NAFLD) is important to prevent progression of the disease. Recently, a low-cost portable magnetic resonance (MR) system was developed as a point-of-care screening tool for in vivo liver fat quantification. However, subcutaneous fat may confound the liver fat quantification, particularly in the NAFLD population. In this work, we propose a novel radiofrequency (RF) coil design composed of a set of "saturation" coils sandwiching a main coil to improve human liver fat quantification. By comparison with conventional MR imaging, we demonstrate the capability and effectiveness of the novel RF coil design in phantom experiments as well as in vivo liver scans. In the phantom experiment, the saturation coil reduced the error in the measured proton density fat fraction (PDFF) results from 28.9% to 4.0%, and in the in vivo experiment, it reduced the discrepancy in the PDFF results from 13.2% to 4.0%. The novel coil design, together with the adapted Carr-Purcell-Meiboom-Gill-based sequence, improves the practicability and robustness of the portable single-side MR system.