Oncology-specific radiation dose and image noise reference levels in adult abdominal-pelvic CT
Moiz Ahmad1, Xinming Liu1, Ajaykumar C Morani2
1Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, 1400 Pressler Street, Unit 1473, Houston, TX 77030-4009, United States of America.
Objectives:
To provide our oncology-specific adult abdominal-pelvic CT reference levels for image noise and radiation dose from a high-volume, oncologic, tertiary referral center.
Methods:
The portal venous phase abdomen-pelvis acquisition was assessed for image noise and radiation dose in 13,320 contrast-enhanced CT examinations. Patient size (effective diameter) and radiation dose (CTDIvol) were recorded using a commercial software system, and image noise (Global Noise metric) was quantified using a custom processing system. The reference level and range for dose and noise were calculated for the full dataset, and for examinations grouped by CT scanner model. Dose and noise reference levels were also calculated for exams grouped by five different patient size categories.
Results:
The noise reference level was 11.25 HU with a reference range of 10.25-12.25 HU. The dose reference level at a median effective diameter of 30.7 cm was 26.7 mGy with a reference range of 19.6-37.0 mGy. Dose increased with patient size; however, image noise remained approximately constant within the noise reference range. The doses were 2.1-2.5 times than the doses in the ACR DIR registry for corresponding patient sizes. The image noise was 0.63-0.75 times the previously published reference level in abdominal-pelvic CT examinations.
Conclusions:
Our oncology-specific abdominal-pelvic CT dose reference levels are higher than in the ACR dose index registry and our oncology-specific image noise reference levels are lower than previously proposed image noise reference levels.
Advances In Knowledge:
This study reports reference image noise and radiation dose levels appropriate for the indication of abdomen-pelvis CT examination for cancer diagnosis and staging. The difference in these reference levels from non-oncology-specific CT examinations highlight a need for indication-specific, dose index and image quality reference registries.
Related Concept Videos
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Imaging Studies I: Kidney, Ureter, and Bladder Studies