Effective clinical applications of Monte Carlo-based independent secondary dose verification software for helical
Takahiro Kosaka1, Jun Takatsu2, Tatsuya Inoue3
1Department of Radiation Oncology, Graduate School of Medicine, Juntendo University, 2-1-1 Hongo, Bunkyo-ku, Tokyo 113-8421, Japan; Department of Radiology, Juntendo University Urayasu Hospital, 2-1-1 Tomioka, Urayasu-shi, Chiba 279-0021, Japan.
Purpose:
To investigate the scope of the effective clinical application of Monte Carlo (MC)-based independent dose verification software for helical tomotherapy.
Methods:
DoseCHECK was selected as the MC-based dose calculation software. First, the dose calculation accuracy of DoseCHECK was evaluated with film and chamber measurements in a water-equivalent phantom. Second, the dose calculation accuracy was examined in several heterogeneous materials. Finally, dosimetric comparisons between DoseCHECK and the treatment planning system (TPS) were performed for clinical patient plans. Prostate IMRT, head and neck IMRT (HN), total body irradiation (TBI), and brain stereotactic radiotherapy (SRT) were evaluated.
Result:
The DoseCHECK calculations agreed with the chamber and film measurements in the homogenous phantom. For heterogeneous phantom cases, the dose differences between DoseCHECK and TPS were within 3 %, except in air, in which large dose differences of 20 % were observed. In clinical patient plans, the median dose differences between the lung Dmean in TBI cases and the normal brain Dmean in brain SRT cases were significantly >3 %. For HN and brain SRT cases, the median target dose differences were >3 %.
Conclusion:
Our results show that independent dose verification with the MC algorithm can detect systematic errors caused by the lack of heterogeneity correction in the TPS. In particular, MC-based independent dose verification is required for HN, TBI, and brain SRT cases in helical tomotherapy.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:25Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiation: Applications
The average...