End-to-end test for lung SBRT: An Italian multicentric pilot experience
S Pallotta1, S Calusi2, L Marrazzo3
1Department of Biomedical, Experimental and Clinical Sciences "Mario Serio", University of Florence, Florence, Italy; Medical Physics Unit, AOU Careggi Florence, Italy.
This study developed and validated an end-to-end (e2e) test for lung stereotactic body radiation therapy (SBRT). Multicenter results confirm the test
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Quality Assurance
Background:
- Stereotactic body radiation therapy (SBRT) for lung cancer requires precise dose delivery.
- End-to-end (e2e) testing is crucial for validating radiotherapy workflows.
- Standardized e2e tests are needed to ensure quality and consistency in lung SBRT across institutions.
Purpose of the Study:
- To establish a comprehensive end-to-end (e2e) testing protocol for lung SBRT.
- To validate the feasibility and effectiveness of this e2e test through a multicenter study.
- To gather user feedback on the test's usability and suitability for clinical practice.
Main Methods:
- A multicenter group of medical physicists designed and implemented an e2e test protocol.
- Four participating centers (PCs) used the ADAM breathing phantom, microDiamond detector, and radiochromic films.
- PCs simulated, planned, and irradiated the phantom according to the protocol, uploading data for centralized analysis.
Main Results:
- All participating centers successfully completed the e2e lung SBRT test in both static and dynamic modes.
- Measured-to-planned dose ratios were generally close to unity across centers and modes.
- Gamma passing rates (3% global, 3mm) demonstrated high accuracy, with most values exceeding 94% in static and dynamic conditions.
Conclusions:
- The proposed end-to-end (e2e) test for lung SBRT is robust and feasible for multicenter validation.
- User feedback indicated the test and protocol were well-suited and clearly described.
- The successful multicenter implementation validates the proposed e2e test for quality assurance in lung SBRT.
Related Concept Videos
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care