Related Experiment Video
Updated: Aug 20, 2025

Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
A new upper limit for effective dose in patient administered with 18F-FDG for PET/CT whole-body imaging with
Parimalah Velo1, Muhammad Irwan Ismail1, Kala Krishnan Mohandas1
1Nuclear Medicine Department, Beacon Hospital Sdn Bhd, Petaling Jaya Selangor, Malaysia.
Introduction:
The aim of present study is to estimate effective dose in patient undergoing 18F-FDG for whole body PET/CT imaging with diagnostic CT parameters and identify the lowest achievable total effective dose.
Method:
A total of 2247 PET/CT patients with normal glucose level underwent 18F-FDG-whole body imaging procedures. The 18F-FDG dose of 3.7MBq per kg of patient weight administered via intravenous infusion. For CT parameters, kilovoltage of 140keV and current of 40 mAs were used for all studies. All the acquired images collected retrospectively and the effective dose was calculated for each patient using algorithm adapted from ICRP Publication 106, modified for patient weight and patient blood volume. The estimated effective doses were evaluated for patients' body weight and BMI.
Results:
The mean of total effective dose and standard deviation is approximately 15.08(4.52) mSv using ICRP algorithm. 56% of total patient has normal BMI and their average total effective dose is 13.6mSv. Underweight patients' effective dose can be as low as 9.6mSv even using diagnostic CT protocols.
Conclusion:
The effective dose of PET/CT procedure in present study is one of the lowest although using diagnostic parameters for CT acquisition compared to published data worldwide. This is due to the improved sensitivity of PET and complex reconstruction technique that maintains the image quality. A significant association between body weight, BMI and effective dose is reported in present study. Therefore, it is suggested that attention must be given for underweight and ideal BMI patients while prescribing FDG activity and CT imaging parameters in order to minimize the effective dose. The effective dose reported in present study can be considered as an upper limit for effective dose in PET/CT patients with normal BMI. This upper limit can be treated as a standard limit when optimizing imaging parameters, developing algorithm for image reconstruction and prescribing activity for patients. This practice could fulfill ALARA principle that could reduce cancer risk.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...