Reproducibility assessment of uptake on dedicated breast PET for noise discrimination
Shunsuke Yuge1, Kanae K Miyake2, Takayoshi Ishimori1
1Department of Diagnostic Imaging and Nuclear Medicine, Graduate School of Medicine, Kyoto University, Kyoto City, Kyoto, Japan.
Objectives:
Dedicated breast PET (dbPET) systems have improved the detection of small breast cancers but have increased false-positive diagnoses due to an increased chance of noise detection. This study examined whether reproducibility assessment using paired images helped to improve noise discrimination and diagnostic performance in dbPET.
Methods:
This study included 21 patients with newly diagnosed breast cancer who underwent [18F]FDG-dbPET and contrast-enhanced breast MRI. A 10-min dbPET data scan was acquired per breast, and two sets of reconstructed images were generated (named dbPET-1 and dbPET-2, respectively), each of which consisted of randomly allocated 5-min data from the 10-min data. Uptake spots higher than the background were indexed for the study with visual assessment. All indexed uptakes on dbPET-1 were evaluated using dbPET-2 for reproducibility. MRI findings based on the Breast Imaging-Reporting and Data System (BI-RADS) 2013 were used as the gold standard. Uptake spots that corresponded to BI-RADS 1 on MRI were considered noise, while those with BI-RADS 4b-6 were considered malignancies. The diagnostic performance of dbPET for malignancy was evaluated using four different criteria: any uptake on dbPET-1 regarded as positive (criterion A), a subjective visual assessment of dbPET-1 (criterion B), reproducibility assessment between dbPET-1 and dbPET-2 (criterion C), and a combination of B and C (criterion D).
Results:
A total of 213 indexed uptake spots were identified on dbPET-1, including 152, 15, 6, 6, and 34 lesions classified as BI-RADS MRI categories 1, 2, 4b, 4c, and 5, respectively. Overall, 31.9% of the index uptake values were reproducible. All malignant lesions were reproducible, whereas 93.4% of noise was not reproducible. The sensitivities for malignancy for criteria A, B, C, and D were 100%, 91.3%, 100%, and 91.3%, respectively, with positive predictive values (PPVs) of 21.4%, 68.9%, 67.6%, and 82.4%, respectively.
Conclusions:
Our results demonstrated that reproducibility assessment helped reduce false-positive findings caused by noise on dbPET without lowering the sensitivity for malignancy. While subjective visual assessment was also efficient in increasing PPV, it occasionally missed malignant uptake.