Hybrid active and passive local shimming (HAPLS) for two-region MRI
Zhi Hua Ren1, Jason Stockmann2,3, Andrew Dewdney4
1Zuckerman Mind Brain Behavior Institute, Columbia University, New York, New York, USA.
Magnetic Resonance in Medicine
|November 28, 2022
Summary
This study introduces a hybrid active and passive shimming technique to simultaneously improve magnetic field homogeneity in two isolated regions of interest within MRI scans. This novel method enhances imaging quality for complex scans like bilateral breasts or dyadic brains.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Biomedical Engineering
- Physics
Background:
- Standard MRI scanners use global shim systems, limiting simultaneous shimming of multiple regions of interest (ROIs).
- Simultaneous shimming of two isolated ROIs is crucial for advanced MRI applications, including bilateral breast imaging and dyadic brain studies.
- Existing technologies struggle to achieve optimal shimming for two distinct areas concurrently.
Purpose of the Study:
- To develop and evaluate a hybrid active and passive local shimming technique for simultaneous shimming of two isolated ROIs.
- To address the limitations of current MRI systems in handling multi-ROI shimming requirements.
- To improve the efficiency and effectiveness of shimming in complex scan scenarios.
Main Methods:
- A local passive shimming system was designed using optimized bilateral ferromagnetic chip arrays to correct high-order B0 inhomogeneities.
- A 40-channel DC loop system powered by 64-channel current amplifiers was employed for active shimming.
- Optimized current distribution was used to correct residual low-order and subject-specific B0 inhomogeneities, and homogenize center frequencies of the two ROIs.
Main Results:
- The hybrid shimming technique reduced 95% peak-to-peak B0 inhomogeneity from 1.92 to 1.12 ppm (41.7% reduction).
- Root Mean Square (RMS) B0 inhomogeneity decreased from 0.473 to 0.255 ppm (46.1% reduction).
- The volume of MR voxels within a 0.5-ppm frequency span increased from 64.3% to 81.3% (26.3% increase).
Conclusions:
- The proposed hybrid active and passive local shimming technique effectively enables simultaneous shimming of two isolated areas.
- This technique addresses an unmet need in commercial MRI scanners for multi-ROI shimming.
- The method significantly improves magnetic field homogeneity, enhancing potential MRI applications.