Objective:
To establish a method for qualitative determination of dichloromethane (DCM) in blood by gas chromatography-mass spectrometry (GC-MS) and quantitative determination of DCM in blood by headspace gas chromatography (HS-GC), and to provide reliable support for forensic examination and analysis of poisoning or deaths caused by DCM.
Methods:
0.5 mL blood sample was collected, added into headspace vial with chloroform as the internal standard, and processed by heating at 65 °C and evacuation treatment. The intermediate gas in the headspace vial was analyzed by GC-MS for qualitative validation of the method and by HS-GC for quantitative validation of the method. The method was then applied in forensic case analysis.
Results:
Qualitative validation of the examination method by GC-MS found that the chromatographic peak and mass spectral characteristic ions were specific in samples added with DCM, and that no interference was observed in the blank negative samples. The limit of detection (LOD) was 5 μg/mL. Quantitative method validation by HS-GC found that the chromatographic peak of DCM was well separated from those of eight other volatile compounds, with the resolution>1.5 in all cases; the lower limit of quantification (LOQ) was 20 μg/mL and good linearity was shown within the range of 20 and 1000 μg/mL, R>0.999; the intra-day test precision and inter-day test precision were good (relative standard deviation, or RSD<15% for both) and test accuracy was high (relative error, or δ<15%). With the method established in the study, DCM was detected successfully in the blood of two fatal cases caused by DCM poisoning, with the blood concentration being 470 μg/mL and 915 μg/mL, respectively.
Conclusion:
This method is shown to be a rapid, stable and accurate approach to the qualitative and quantitative forensic and toxicological analysis of DCM in blood in DCM poisoning cases or deaths caused by DCM.
More Related Videos
10:11Microdialysis of Ethanol During Operant Ethanol Self-administration and Ethanol Determination by Gas Chromatography
Published on: September 5, 2012
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II