Ultrasound insonation angle and scanning imaging modes for imaging dental implant structures: A benchtop study
Oliver D Kripfgans1, Nikhila Devi Goli2, Jad Majzoub2
1Department of Radiology, Michigan Medicine, University of Michigan, Ann Arbor, MI, United States of America.
Introduction:
High frequency ultrasound has shown as a promising imaging modality to evaluate peri-implant tissues. It is not known if the ultrasound imaging settings might influence ultrasound's ability to differentiate implant structures. The aim of this benchtop study was to evaluate the dependence of ultrasound on imaging angles and modes to measure implant geometry-related parameters.
Methods:
A clinical ultrasound scanner (ZS3, Mindray) with an intraoral probe (L30-8) offering combinations of harmonic and compound imaging modes was employed for imaging 16 abutments and 4 implants. The samples were mounted to a micro-positioning system in a water tank, which allowed a range of -30 to 30-degree imaging angles in 5-degree increment between the probe and samples. The abutment angle, implant thread pitch and depth were measured on ultrasound, compared to the reference readings. The errors were computed as a function of the image angles and modes. All samples were replicated 3 times for 3 image modes and 11 image angles, thus resulting in 2,340 images.
Results:
The mean errors of ultrasound to estimate 16 abutment angles, compared to the reference values, were between -1.8 to 2.7 degrees. The root mean squared error (RMSE) ranged from 1.5 to 4.6 degrees. Ultrasound significantly overestimated the thread pitch by 26.1 μm to 36.2 μm. The error in thread depth measurements were in a range of -50.5 μm to 39.6 μm, respectively. The RMSE of thread pitch and depth of the tested 4 implants was in a range of 34.7 to 56.9 μm and 51.0 to 101.8 μm, respectively. In most samples, these errors were independent of the image angle and modes.
Conclusions:
Within the limitations of this study, high-frequency ultrasound was feasible in imaging abutments and implant fixtures independent of scanning angle within ±30° of normal incidence and for compounding and non-compounding-based imaging modes.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies II: Ultrasonography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...