Efficacy and safety of ultrasound cycloplasty in Indian eyes with open-angle glaucoma
Chandrima Paul1, J Divya2, Subhrangshu Sengupta3
1Director, B B Eye Foundation, Kolkata, West Bengal, India.
Purpose:
To evaluate the safety and efficacy of ultrasound cycloplasty in eyes with primary or secondary open-angle glaucoma, not amenable to adequate control of intra-ocular pressure (IOP) with medical treatment.
Methods:
Prospective interventional cohort study of 28 eyes of 28 subjects in a tertiary eye care centre in India in patients with open-angle glaucoma. All enrolled eyes underwent ultrasound cycloplasty with the second-generation probe with six shots of 8 s each, operated by a single surgeon between November 2018 and January 2020. They were followed up for a period of 12 months. The primary treatment outcome was IOP and the secondary outcomes were vision and postoperative complications.
Results:
A total of 28 eyes of 28 patients were studied, and the mean age was 63.82 ± 6.46 years. Primary open-angle glaucoma (75%) was the most common etiology. There was significant reduction in IOP from the baseline (24.93 ± 4.27 mmHg) to the postoperative value (15.82 ± 3.14 mmHg) at the end of 12 months (P < 0.00001). Mean reduction in IOP was 9.14 ± 4.09 mmHg at 12 months (36.66%). Number of ocular hypotensives reduced significantly from baseline (3.32 ± 0.47) to 12-month postoperative follow-up (0.68 ± 0.74) (P < 0.00001). Qualified success was achieved in 89.28% eyes. No major complications were noted.
Conclusion:
Ultrasound cycloplasty is found to be effective and safe in eyes with open-angle glaucoma because of the primary or secondary etiology, being more effective in the former.