Comparison of peripheral extension, acquisition time, and image chromaticity of Optos, Clarus, and EIDON systems
Federico Fantaguzzi1,2, Andrea Servillo1,2, Riccardo Sacconi1,2
1School of Medicine, Vita-Salute San Raffaele University, Via Olgettina 60, 20132, Milan, Italy.
Purpose:
To evaluate differences in acquisition time, peripheral extension, and chromaticity between 3 different commercialized ultra-wide-field (UWF) fundus cameras.
Methods:
Patients were prospectively enrolled from 07/2021 to 11/2021. Patients underwent fundus photography with the following scanning protocols: (1) single shot with Silverstone (Optos, California), two-shot montage with Clarus 500 (Carl Zeiss, Dublin, CA), and three-shot montage with iCare EIDON FA with UWF module (CenterVue Spa, a company of iCare Finland Oy; Vantaa, Finland). Acquisition time was calculated as the interval between the beginning and the end of the acquisition. Peripheral extension was quantified as the average ratio between the total retinal pixel area and the optic nerve head (ONH) pixel area. The average chromaticity of all pixels in the red-green-blue (RGB) space was calculated.
Results:
Twenty-three eyes of 13 prospectively enrolled healthy controls were included in the study. Optos Silverstone had a higher total retina area/ONH area ratio (509.1 [480.9;559.3]) compared to Zeiss Clarus (442.0 [431.9;510.5], p = 0.02) and iCare EIDON (369.7 [345.3;387.8], p < 0.0001). Silverstone demonstrated the shortest acquisition time (median [interquartile range]: 32 [20;58.5] s) compared to Zeiss Clarus (42 [28.5;53.5] s, p = 0.6733) and iCare EIDON (72 [68.5;78] s, p = 0.0003). iCare EIDON demonstrated the lowest variability of acquisition time (9.5 s), compared to Zeiss Clarus (25 s) and Optos Silverstone (38.5 s). A statistically significant difference was found in the RGB distribution between each of the 3 devices (p < 0.001). iCare EIDON demonstrated an average barycenter position (RGB = [0.412, 0.314, 0.275]) that represented the best color balance of the image. Zeiss Clarus had a noticeable red shift at the expense of the blue and green channels (RGB = [0.515, 0.294, 0.191]). Optos Silverstone showed an absence of the blue channel (RGB = [0.621, 0.372, 0.007]) which results in a distortion of the color of the image.
Conclusion:
Optos Silverstone and Zeiss Clarus required less time than iCare EIDON to acquire a comparable size image and captured larger areas of the retina than iCare EIDON. iCare EIDON provided more color-balanced retinal images with greater richness of color content than the other two devices.
More Related Videos
07:11Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
06:19In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...