Development and validation of a LC-MS/MS method for quantitation of recombinant human growth hormone in rat plasma
Wen-Si Huang1, Jing-Hua Yu2, Xing-Xing Diao1
1Shanghai Institute of Materia Medica, Chinese Academy of Sciences, Shanghai 201210, China; University of Chinese Academy of Sciences, Beijing 100049, China.
Insights
A new LC-MS/MS method accurately quantifies recombinant human growth hormone (rhGH) in rat plasma. This sensitive assay supports pharmacokinetic studies, offering a faster, simpler alternative for rhGH analysis.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Pharmacology
Background:
- Recombinant human growth hormone (rhGH) is a 191-amino acid peptide crucial for pediatric growth and possesses anti-aging properties.
- Accurate quantification of rhGH in biological matrices is essential for pharmacokinetic and therapeutic monitoring.
Purpose of the Study:
- To develop and validate a simple, sensitive, and rapid LC-MS/MS method for quantifying rhGH in rat plasma.
- To assess the method's suitability for pharmacokinetic studies.
Main Methods:
- Establishment of a quantitative LC-MS/MS assay using enzymatic digestion of plasma samples.
- Selection of two surrogate peptides (LFDNAMLR and FPTIPLSR) for targeted analysis.
- Validation of linearity, accuracy, and precision across a wide calibration range.
Main Results:
- The method demonstrated good linearity from 10-2000 ng/mL.
- Quality control accuracy ranged from -13.8% to 14.3%, with LLOQ accuracy between -12.9% and 19.0%.
- Intra-day and inter-day precision were within acceptable limits (1.7-13.6% and 4.0-7.0%, respectively).
Conclusions:
- The developed LC-MS/MS method is simple, sensitive, and efficient for rhGH quantification in rat plasma.
- The method's advantages include a short processing time (3.5 h), wide linear range, small plasma volume requirement (35 μL), and a low LLOQ (10 ng/mL).
- This validated assay is suitable for intravenous and subcutaneous pharmacokinetic studies of rhGH in rats.
Abstract:
Recombinant human growth hormone (rhGH) is a peptide comprising 191 amino acids, that is mainly used to promote the growth of children and plays an important antiaging role. In the present study, a simple and sensitive quantitation method for rhGH in rat plasma was established by LCMS/MS. After simple and rapid enzymatic digestion of the plasma sample, two suitable surrogate peptides (LFDNAMLR and FPTIPLSR) were selected for quantitative analysis. The results showed good linearity over calibration range 10-2000 ng/mL. The quality control (QC) accuracy ranged from -13.8 to 14.3%, and the accuracy of the lower limit of quantification (LLOQ) ranged from -12.9 to 19.0%. The intra-day and inter-day precision ranges for all QCs were 1.7-13.6% and 4.0-7.0%, respectively. The method was successfully applied to intravenous and subcutaneous pharmacokinetic studies in rats. In comparison with previously published methods, our method features simple sample preparation combined with a short sample processing time (3.5 h), wide linear range (10-2000 ng/mL), small plasma volume (35 μL), and LLOQ (10 ng/mL).
More Related Videos
11:36A RAPID Method for Blood Processing to Increase the Yield of Plasma Peptide Levels in Human Blood
Published on: April 28, 2016
11:09An HS-MRM Assay for the Quantification of Host-cell Proteins in Protein Biopharmaceuticals by Liquid Chromatography Ion Mobility QTOF Mass Spectrometry
Published on: April 17, 2018