Correlation between osmolarity measurements using the TearLab™ and I-Pen® systems in subjects with a high body mass
Mana A Alanazi1, Gamal A El-Hiti1, Naif Alhafy1
1Optometry Department, College of Applied Medical Sciences, King Saud University, Riyadh, Saudi Arabia.
Background:
Osmolarity is used to detect symptoms of dry eye disease (DED) and can be measured using TearLab™ and I-Pen® systems.
Objectives:
To investigate the correlation between osmolarity measurements using the TearLab™ and I-Pen® systems in subjects with a high body mass index (BMI).
Material And Methods:
Thirty male subjects with a high BMI (27-48 kg/m2; 23.3 ±2.1 years old) participated in this study. The control group consisting of 30 healthy males (24.9 kg/m2; 22.9 ±2.1 years old) was also enrolled. Osmolarity measurements were recorded from the right eye using the TearLab™ and I-Pen® systems, and interviews were conducted to determine ocular surface disease index (OSDI) scores.
Results:
The OSDI (p = 0.042), TearLab™ (p < 0.001) and I-Pen® (p < 0.001) scores were significantly higher in the study group than in the control group. In the study group, OSDI scores ranged from 2 to 16 (median 8.0, interquartile range (IQR) 6.8), while it was from 0 to 10 (median 6.3, IQR 4.1) in the control group. The TearLab™ osmolarity scores were in the range of 278-309 mOsm/L in the study group, whereas the I-Pen® osmolarity measurements were in the range of 294-336 mOsm/L in the study group, compared with 263-304 mOsm/L and 278-317 mOsm/L in the control group, respectively. In the study group, there was a strong correlation between the TearLab™ and I-Pen® osmolarity scores (r = 0.934; p = 0.001). In addition, strong correlations were found between the BMI and both TearLab™ (r = 0.736; p = 0.001) and I-Pen® (r = 0.707; p = 0.001) scores, as well as between the OSDI scores and both TearLab™ (r = 0.731; p = 0.001) and I-Pen® measurements (r = 0.666; p = 0.001).
Conclusion:
Osmolarity measurements using the I-Pen® system were significantly higher than those recorded using the TearLab™ system in subjects with a high BMI. The I-Pen® measurements showed large variations in osmolarity scores and were highly unreliable in correctly identifying normal eyes compared to the TearLab™ system. Also, a strong correlation was found between the osmolarity scores obtained from the TearLab™ and I-Pen® systems.
More Related Videos
07:44Author Spotlight: Implementation of BIVA for Analyzing Disease Risk Factors in Patients with Low Body Cell Mass
Published on: July 14, 2023
03:35Determining Gender-Based Differences in Retinal and Choroidal Thickness in Underweight Individuals via Swept-Source Optical Coherence Tomography
Published on: December 1, 2023
Related Concept Videos
Osmosis and Osmotic Pressure of Solutions
Composition of Body Fluids