Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Validation of Wearable Device Consisting of a Smart Shirt with Built-In Bioelectrodes and a Wireless Transmitter for
Yuki Hashimoto1, Rieko Sato1, Kazuhiko Takagahara1
1NTT Device Innovation Center, NTT Corporation, 3-1, Atsugi 243-0198, Japan.
Abstract:
Real-time monitoring of heart rate is useful for monitoring workers. Wearable heart rate monitors worn on the upper body are less susceptible to artefacts caused by arm and wrist movements than popular wristband-type sensors using the photoplethysmography method. Therefore, they are considered suitable for stable and accurate measurement for various movements. In this study, we conducted an experiment to verify the accuracy of our developed and commercially available wearable heart rate monitor consisting of a smart shirt with bioelectrodes and a transmitter, assuming a real-world work environment with physical loads. An exercise protocol was designed to light to moderate intensity according to international standards because no standard exercise protocol for the validation simulating these works has been reported. This protocol includes worker-specific movements such as applying external vibration and lifting and lowering loads. In the experiment, we simultaneously measured the instantaneous heart rate with the above wearable device and a Holter monitor as a reference to evaluate mean absolute percentage error (MAPE). The MAPE was 0.92% or less for all exercise protocols conducted. This value indicates that the accuracy of the wearable device is high enough for use in real-world cases of physical load in light to moderate intensity tasks such as those in our experimental protocol. In addition, the experimental protocol and measurement data devised in this study can be used as a benchmark for other wearable heart rate monitors for use for similar purposes.
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes