Related Experiment Video
Updated: Aug 17, 2025

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Stability of implant-abutment connection in three different systems after fatigue test
Farshad Bajoghli1, Mohmoud Sabouhi2, Mohamad Pourali3
1Dental Implants Research Center, Department of Prosthodontics, Dental Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.
Aim:
Abutment screw loosening of implant-supported prosthesis causes a mismatch between the abutment and the implant. This screw loosening is influenced by the implant-abutment connection type, however, with contradictory results reported in different studies. The present study evaluates the stability of abutment-implant connections in three different systems before and after the fatigue test.
Settings And Design:
Thirty implants (4.3 mm in diameter and 12 mm in length) were divided into three groups of 10: Implantium, Zimmer, and straight internal hexagonal connection (SIC) implants.
Materials And Methods:
Two torques of 35 Ncm with an interval of 10 min were applied, followed by measuring removal torque value (RTV). The samples were re-torqued and then underwent a simulation of 1-year chewing clinical performance of dental implant under axial force of 400 N, with a frequency of 8 Hz (one million cycles). After fatigue test, the RTV was calculated and recorded.
Statistical Analysis:
The mean RTVs obtained before and after cyclic load were analyzed by SPSS version 22 software using multivariate analysis.
Results:
Significant differences in RTV and role of cyclic loading were found between SIC and Implantium groups (P = 0.006 and 0.021, respectively), as well as between Zimmer and SIC groups (P = 0.032 and 0.006, respectively), but not between Zimmer and Implantium groups (P = 0.771 and 0.248, respectively).
Conclusion:
The type of connection could affect the screw loosening, the preload loss, and the implant component stability. SIC group revealed the highest RTVs before and after cyclic loading.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Related Concept Videos
Fatigue
Stability of structures
Applications of Stress
The...
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...