Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Quantification and dosimetry of small volumes including associated uncertainty estimation
Lily Carnegie-Peake1,2, Jan Taprogge1,2, Iain Murray1,2
1Joint Department of Physics, Royal Marsden NHSFT, Downs Road, Sutton, SM2 5PT, UK.
Background:
Accurate quantification of radioactivity in a source of interest relies on accurate registration between SPECT and anatomical images, and appropriate correction of partial volume effects (PVEs). For small volumes, exact registration between the two imaging modalities and recovery factors used to correct for PVE are unreliable. There is currently no guidance relating to quantification or the associated uncertainty estimation for small volumes.
Material And Methods:
A method for quantification of small sources of interest is proposed, which uses multiple oversized volumes of interest. The method was applied to three Na[131I]I activity distributions where a Na[131I]I capsule was situated within a cylindrical phantom containing either zero background, uniform background or non-uniform background and to a scenario with small lesions placed in an anthropomorphic phantom. The Na[131I]I capsule and lesions were quantified using the proposed method and compared with measurements made using two alternative quantification methods. The proposed method was also applied to assess the absorbed dose delivered to a bone metastasis following [131I]mIBG therapy for neuroblastoma including the associated uncertainty estimation.
Results:
The method is accurate across a range of activities and in varied radioactivity distributions. Median percentage errors using the proposed method in no background, uniform backgrounds and non-uniform backgrounds were - 0.4%, - 0.3% and 1.7% with median associated uncertainties of 1.4%, 1.4% and 1.6%, respectively. The technique is more accurate and robust when compared to currently available alternative methods.
Conclusions:
The proposed method provides a reliable and accurate method for quantification of sources of interest, which are less than three times the spatial resolution of the imaging system. The method may be of use in absorbed dose calculation in cases of bone metastasis, lung metastasis or thyroid remnants.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Uncertainty: Overview
Uncertainty in Measurement: Reading Instruments
Estimation of the Physical Quantities
Uncertainty in Measurement: Accuracy and Precision
Propagation of Uncertainty from Systematic Error
Uncertainty: Confidence Intervals