Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Calibration frequency and analytical variability of laboratory measurements
Chun Yee Lim1, Shin Ow Yang1, Corey Markus2
1Engineering Cluster, Singapore Institute of Technology, Singapore.
Background:
There is uncertainty whether increased frequency of calibrations may affect the overall analytical variability of a measurement procedure as reflected in quality control (QC) performance. In this simulation study, we examined the impact of calibration frequencies on the variability of laboratory measurements.
Methods:
A 5-point calibration curve was modeled with simulated concentrations ranging from 10 to 10,000 mmol/l, and signal intensities with CVs of 3 % around the mean, under a Gaussian distribution. 3 levels of QC (20, 150, 600 mmol/l) interspersed within the analytical measurement range were also simulated.
Results:
The CV of the 3 QC levels remained stable across the different calibration frequencies simulated (5, 10, 15 and 30 QC measurements per recalibration episode). The imprecision was greatest (18 %) at the lowest concentration of 20 mmol/l, when the calibration curve was derived using ordinary least squares regression, reducing to 3.5 % and 3.8 % at 150 and 600 mmol/l, respectively. The CV of all 3 QC concentrations remained constant at 3.4 % and close the predefined CV (3 %) when weighted least squares regression was used to derive the calibration model. Similar findings were observed with 2-point calibrations using WLS models at narrower concentration ranges (50 and 100 mmol/l as well as 50 and 500 mmol/l).
Discussion:
Within the parameters of the simulation study, an increased frequency of calibration events does not adversely impact the overall analytical performance of a measurement procedure under most circumstances.
More Related Videos
10:00Calibration of Vector Network Analyzer for Measurements in Radio Frequency Propagation Channels
Published on: June 2, 2020
08:52Improving Reproducibility to Meet Minimal Information for Studies of Extracellular Vesicles 2018 Guidelines in Nanoparticle Tracking Analysis
Published on: November 17, 2021
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Random and Systematic Errors
Uncertainty in Measurement: Accuracy and Precision
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Data Validation
Key parameters for method validation include:
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...