Related Experiment Video
Updated: Jun 23, 2026

Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Preclinical Studies of the Nebulized Delivery of Liposomal Amphotericin B
Sagar Kothari1, Steven G Kefalos2, Nicholas D Hages2
1Division of Infectious Diseases, Ajmera Transplant Center, University Health Network, Toronto, Ontario, Canada.
Abstract:
Intravenous liposomal amphotericin B (L-AMB) has accompanying side effects that may be diminished when administering an inhaled form. Delivery systems for inhaled or aerosolized L-AMB vary, and there has not been a recent comparison of available systems to date. We compared three differently designed nebulizer delivery systems for the inhaled delivery of L-AMB to determine the best combination of efficient lung dosing and treatment time. Aerosol size was measured using a Malvern Mastersizer, and five separate nebulizers were tested. For drug output measurements, a Harvard Lung was used, and aerosol was collected using HEPA filters. Overall aerosol size characteristics were similar for all devices with volume median diameters in the 4-5 μm range. The highest inhaled dose was delivered by the AeroEclipse. The Aerogen and the AeroEclipse had similar predicted pulmonary doses, and the AeroEclipse had the highest pulmonary delivery rates. The AeroEclipse nebulizer may provide more efficient delivery in a shorter amount of time; however, human studies are warranted to assess the safety, tolerability, and efficacy of inhaled delivery of L-AMB from this system.