Neck Strength Evaluated With Fixed and Portable Dynamometers in Asymptomatic Individuals: Correlation, Concurrent
Tais S Martins1, Carina F Pinheiro-Araujo1, Camila Gorla1
1Department of Health Sciences, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil.
Fixed and portable dynamometers show moderate to high correlation for neck strength but do not agree on measurements. For consistent neck strength assessment, always use the same dynamometer equipment.
Area of Science:
- Biomechanics
- Musculoskeletal assessment
- Rehabilitation engineering
Background:
- Accurate measurement of cervical isometric force is crucial for assessing neck strength and guiding rehabilitation.
- Both fixed and portable dynamometers are used, but their comparability in measuring cervical isometric force is not well-established.
Purpose of the Study:
- To evaluate the correlation, concurrent validity, and agreement between fixed and portable dynamometers for isometric cervical force measurements in asymptomatic individuals.
Main Methods:
- Fifty asymptomatic individuals performed maximal isometric cervical contractions (flexion, extension, lateral flexion) using both fixed and portable dynamometers.
- Spearman's correlation coefficient and intraclass correlation coefficient (ICC) were used to assess correlation and validity.
- Bland-Altman analysis was employed to evaluate agreement between the two devices.
Main Results:
- A moderate to high correlation (rs = 0.68-0.82) and moderate to good concurrent validity (ICC = 0.67-0.80) were found between the dynamometers for all movements.
- However, significant mean differences and broad limits of agreement were observed, indicating a lack of measurement equivalence.
- Errors ranged from 61% to 77% of the mean force measured by the fixed dynamometer.
Conclusions:
- While fixed and portable dynamometers show good correlation and validity for cervical isometric force, they do not provide interchangeable measurements.
- Due to significant discrepancies, it is recommended to consistently use the same type of dynamometer for serial reassessments of neck strength.
More Related Videos
07:30Muscle Imbalances: Testing and Training Functional Eccentric Hamstring Strength in Athletic Populations
Published on: May 1, 2018
12:59Improving Strength, Power, Muscle Aerobic Capacity, and Glucose Tolerance through Short-term Progressive Strength Training Among Elderly People
Published on: July 5, 2017
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...