Related Experiment Video
Updated: Aug 17, 2025

Measurement of Maximum Isometric Force Generated by Permeabilized Skeletal Muscle Fibers
Published on: June 16, 2015
Portable Dynamometer-Based Measurement of Pelvic Floor Muscle Force
Batoul El-Sayegh1,2, Chantale Dumoulin2,3, Mohamed Ali1,4
1Department of Electrical EngineeringPolytechnique Montreal Montreal QC H3T 1J4 Canada.
Objective:
In attempts to improve the quality of life of women, continuous projects are sought between rehabilitation intervention and engineering. Using the knowledge of the pelvic floor muscle (PFM) physiology, assessment and training methods are developed to reduce lower urinary tract symptoms such as urinary incontinence. Therefore, this paper covers the design and implementation of a portable vaginal dynamometer.
Methods:
A PFM probe is designed, 3D printed, assembled, and tested in ten women to assess its acceptability and usability. The feedback from the usability study is used to optimize the PFM probe design. A vaginal dynamometer is developed based on the designed PFM probe, then tested for linearity, repeatability, hysteresis, noise and heat effect, and power consumption. The variability between the different produced PFM probe prototypes is evaluated.
Results:
Force measurements are made using a load cell. Wireless communication is performed through a Bluetooth low energy transceiver v5.0, with a corresponding interface on both computer and smartphone. The device operates at a 3.3V supply and achieves a power consumption of 49.5 mW in operating mode. Two PFM probe sizes are designed to accommodate different vaginal hiatus sizes, based on usability study feedback. The proposed system allows the physiotherapist to wirelessly monitor variation in pelvic floor muscle force during assessment and/or training.
Discussion/Conclusion:
The testing results showed that the newly designed system has the potential to measure the PFM function in functional conditions such as the standing position.