Related Experiment Video
Updated: Aug 17, 2025

In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
Published on: July 2, 2021
Posterior-stabilized total knee arthroplasty kinematics and joint laxity: A hybrid biomechanical study
Allan R Sekeitto1, Jance G McGale2, Liam A Montgomery3
1London Health Sciences Centre, 339 Windermere Rd, London, ON, N6A 5A5, Canada. sekeitto@yahoo.com.
Background:
Posterior-stabilized (PS)-total knee arthroplasty (TKA) arose as an alternative to cruciate-retaining (CR)-TKA in the 1970s. Since then, it has become a popularly utilized TKA design with outcomes comparable to CR-TKA. The post-cam mechanism is unique to PS-TKA as it substitutes the function of the posterior cruciate ligament (PCL). The study aimed to understand the kinematic and laxity changes in PS-TKA with under- and overstuffing of the tibiofemoral joint space with the polyethylene (PE) insert.
Methods:
This study employed a hybrid computational-experimental joint motion simulation on a VIVO 6 degrees of freedom (6-DoF) joint motion simulator (AMTI, Watertown, MA, USA). Physical prototypes of a virtually-performed TKA in mechanical alignment (MA) and kinematic alignment (KA) based on cadaveric CT scans and a virtual ligament model were utilized. The reference, understuffed (down 2 mm) and overstuffed (up 2 mm) joint spaces were simulated, neutral flexion and laxity testing loads and motions were performed for each configuration.
Results:
The PE insert thickness influenced post-cam engagement, which occurred after 60º in the overstuffed configurations, after 60º-75º in the reference configurations and after 75º in the understuffed configurations. The understuffed configurations, compared to the reference configurations, resulted in a mean 2.0º (28%) and 2.0º (31%) increase in the coronal laxity in MA and KA respectively. The overstuffed configurations, compared to the reference configuration, resulted in an increase in the mean joint compressive forces (JCFs) by 73 N (61%) and 77 N (62%) in MA and KA models, respectively.
Conclusions:
The under- and overstuffing in PS-TKA alter the kinematics with variable effects. Understuffing decreases the stability, JCFs and inverse with overstuffing. Subtle changes in the PE insert thickness alter the post-cam mechanics.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Development of the Limb Synovial Joints
The mesenchymal stem cells differentiate into chondrocytes that form the hyaline cartilage, and later the cartilaginous model of the bone. This model further transforms into a bone. This process is known as endochondral ossification.
During development, the limbs...