Unexpected dosing errors due to air bubbles in infusion lines with and without air filters
Maurits K Konings1, Kelly Haaijer1, Robin Gevers1
1Dept. of Medical Technology and Clinical Physics, University Medical Center Utrecht, Utrecht, Netherlands.
Abstract:
The effect of the presence of an air bubble, inside an infusion line, on the time (Tnew) needed for a new medication to reach the patient after a syringe exchange was studied in this paper. If an air bubble escapes through an air filter, then a sudden drop in pressure occurs, causing a relaxation of the compressible part of the syringe, followed by a gradual restoration of the flow rate in the line. We modeled this phenomenon mathematically and measured it experimentally in vitro. In an example with a pump flow rate of 5 mL/h and an air bubble of 1 cm length inside an infusion line (diameter 1 mm) with an air filter, both theory and experiment yield an additional increase of at least 600% in delay time if a naive estimate (based on the size of the bubble alone) is replaced by a more realistic estimate incorporating compressibility. Furthermore, we show that an air bubble in a line without air filter may increase Tnew by a factor 2, depending on the initial position of the air bubble. We conclude that an air bubble in an infusion line causes delays that may not be expected by health care professionals.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Inhaled Medications
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...