Verification of patient-setup accuracy using a surface imaging system with steep measurement angle
Fumihiro Sasaki1, Yuusuke Yamashita2, Satoshi Nakano1
1Teine Keijinkai Hospital, Sapporo, Hokkaido, Japan.
Journal of Applied Clinical Medical Physics
|December 20, 2022
Summary
Surface setup (SS) using Voxelan at a 74-degree angle demonstrated accuracy comparable to laser setup (LS) for patient positioning. Phantom tests and patient data showed similar inter-fractional setup errors, indicating Voxelan
Area of Science:
- Medical Physics
- Radiation Oncology
- Image Guidance
Background:
- Accurate patient positioning is critical in radiation therapy.
- Surface-guided radiation therapy (SGRT) offers a non-contact method for patient setup.
- Evaluating SGRT systems at non-standard installation angles is important for broader clinical adoption.
Purpose of the Study:
- To assess the accuracy of the Voxelan HEV-600 M/RMS SGRT device installed at a steep 74-degree camera view angle with the Radixact system.
- To compare the accuracy of Voxelan's surface setup (SS) against traditional laser setup (LS) using phantom experiments and patient data.
Main Methods:
- Phantom experiments were conducted to evaluate Voxelan's accuracy in translations and rotations against ground truth.
- Inter-fractional setup errors were compared between laser setup (LS) and surface setup (SS) in head and neck, chest, and pelvis patient cases.
- Setup errors were determined by comparing surface setup to bone matching using megavoltage computed tomography (MVCT) as the reference standard.
Main Results:
- Phantom experiments showed Voxelan's measured values to be within 1 mm and 1 degree of the true values.
- No significant difference in inter-fractional setup error was found between LS and SS across all patient cohorts (P > 0.05).
- Slight improvements in setup accuracy were noted for head and neck and chest cases with SS, while pelvis cases showed no improvement due to visualization challenges.
Conclusions:
- Surface setup using Voxelan at a 74-degree installation angle achieves accuracy comparable to or better than traditional laser setup.
- The Voxelan system, even at a steep angle, provides reliable patient positioning for radiation therapy.
- Further investigation may be needed for specific regions like the pelvis where surface visualization can be challenging.