Lead-Dust Contamination on Radiation Protection Apparel
Felicia Manocchio1, Tiffany Ni2, Gaylene Pron3
1School of Public Health, University of Waterloo, Waterloo, Ontario, Canada.
Purpose:
To evaluate the prevalence of surface lead-dust contamination on radiation protection apparel (RPAs) in the radiology department and compare findings with those from other studies of RPA lead-dust contamination.
Materials And Methods:
A survey of RPAs was conducted between June and December 2021 in radiology departments at a tertiary-care university hospital. A convenience sample of RPAs located on wall-mounted racks outside the angiography suite and emergency department was surveyed. Surface lead dust on RPAs was detected using a rapid qualitative test.
Results:
A total of 69 RPAs included full-length frontal lead aprons (n = 11), full-length frontal lead aprons (n = 25) with thyroid collars (n = 25), and thyroid collars alone (n = 8). Garments consisted mainly of a lead/antimony composite core with a 0.5-mm lead equivalency. One RPA failed radiologic quality inspection, and 8 garments were in poor or worn condition. The overall prevalence of surface lead-dust contamination on RPAs was 60.9% (95% CI, 49.1%-71.5%) and was significantly (P = .0035) higher on thyroid collars (78.8% [95% CI, 62.2%-89.3%]) than on lead aprons (44.4% [95% CI, 29.5%-60.4%]).
Conclusions:
A high prevalence of surface lead-dust contamination was detected on RPAs using a rapid qualitative test. There is currently no established safe level of lead, and these findings suggest RPAs be monitored frequently not only for physical defects limiting radiation protection but also for lead-dust contamination.
More Related Videos
09:29An Air-liquid Interface Bronchial Epithelial Model for Realistic, Repeated Inhalation Exposure to Airborne Particles for Toxicity Testing
Published on: May 13, 2020
07:54Experimental Methods of Dust Charging and Mobilization on Surfaces with Exposure to Ultraviolet Radiation or Plasmas
Published on: April 3, 2018
Related Concept Videos
PPE Use in Healthcare Settings II: Doffing
PPE Use in Healthcare Settings I: Donning
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Personal Protective Equipment
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force...
Radiation: Applications
The average...