Related Experiment Video
Updated: Aug 16, 2025

Endotracheal Intubation via Tracheotomy and Subsequent Thoracotomy in Rats for Non-Survival Applications
Published on: March 15, 2024
Tracheal airway pressure in tracheostomy tube capping trials: an experimental study
Andreas Nowak1, Sten Martin2, Maik Höhne2
1Department of Anesthesiology and Intensive Care Medicine, Emergency Medicine and Pain Management, Dresden Friedrichstadt Hospital, Technical University Dresden Teaching Hospital, Friedrichstrasse 41, 01067, Dresden, Germany. andreas.nowak@klinikum-dresden.de.
Background:
Tracheostomy tube capping is a commonly used test to determine if the tracheostomy tube can be removed. The success of the capping trial depends on the patient's ability to maintain sufficient spontaneous breathing with an occluded tracheostomy tube. The impact of an occluded tracheotomy tube on airway resistance is currently unknown. The aim of this study was to investigate tracheal pressure during capping or stoma button insertion and potential determinants concerning cuff.
Methods:
Eight cuffed and uncuffed tracheostomy tubes and three stoma buttons of various manufacturers and sizes were inserted into the trachea model. Cuffs were completely deflated or contained atmospheric pressure. The trachea was ventilated bidirectional with a respirator in volume-controlled mode and volume flows 15-60 L/min. Tracheal pressure drop during inspiration as a parameter of pressure required to move gas through the airway was measured.
Results:
Tracheal pressure drops occurred linearly or irregularly during capping trials to a maximum of 4.2 kPa at flow rates of 60 L/min for atmospheric pressure cuffs. In tracheostomy tubes with completely deflated cuffs, pressure drop in the trachea reaches a maximum of 3.4 kPa at a flow rate of 60 L/min. For tracheostomy tubes with cuff smaller inner or outer diameters do not regularly result in lower tracheal pressure drop. The pressure drop varies between different tracheostomy tubes depending on the manufacturer. In cuffed tracheostomy tubes, we observed three phenomena: sail-like positioning, folding over, and tightening of the cuff during flow. The maximum tracheal pressure drop during stoma button insertion reaches 0.014 kPa.
Conclusions:
The cuff is a central element for the pressure drop in the airway and thus airway resistance during spontaneous translaryngeal breathing with a capped TT. Complete deflation reduces the pressure drop in the trachea. Due to deformation of the cuff, measured pressures are irregular as the volume flow is increased. Incomplete deflated cuffs and material characteristics of tracheostomy tubes and cuffs in addition to anatomical and clinical variables may cause unsuccessful capping trials due to increased airway resistance. All stoma buttons showed that pressure drop and thus airway resistance due to stoma buttons has no clinical relevance.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...