Screw Pullout Strength After Pedicle Screw Reposition: A Finite Element Analysis
Yu-Xing Ye1,2, Da-Geng Huang1,2, Ding-Jun Hao1,2
1Department of Spine Surgery, Honghui Hospital, Xi'an Jiaotong University, Xi'an, Shaanxi, China.
Research Design:
Finite element analysis based on computed tomography images from the lumbar spine.
Objective:
Determined the pullout strength of unsatisfactorily placed screws and repositioned screws after unsatisfactory place in lumbar spine surgery.
Background:
Pedicle screws are widely used to stabilize the spinal vertebral body. Unsatisfactory screws could lead to surgical complications, and may need to be repositioned. Screw removal and reposition, however, may decrease pullout strength.
Methods:
We conducted a three-dimensional finite element analysis based on high-resolution computed tomography images from a 39-year-old healthy woman. Pullout strength was determined with the screw placed in different orientations at the same entry point (as selected by the Magerl method), as well as after removal and reposition. The material properties of the vertebral body and the screw were simulated by using grayscale values and verified data, respectively. A load along the screw axis was applied to the end of the screw to simulate the pullout.
Results:
The pullout strength was 1840.0 N with the Magerl method. For unsatisfactorily placed screws, the pullout strength was 1500.8 N at 20% overlap, 1609.6 N at 40% overlap, 1628.9 N at 60% overlap, and 1734.7 N at 80% overlap with the hypothetical screw path of the Magerl method. For repositioned screws, the pullout strength was 1763.6 N, with 20% overlap, 1728.3 N at 40% overlap, 1544.0 N at 60% overlap, and 1491.1 N at 80% overlap, with the original path. Comparison of repositioned screw with unsatisfactorily placed screw showed 14.04% decrease in pullout strength at 80% overlap, 5.21% decrease at 60% overlap, 7.37% increase at 40% overlap, and 17.51% increase at 20% overlap, with the screw path of the Magerl method.
Conclusions:
Removal and reposition increased the pullout strength at 20% and 40% overlap, but decreased the pullout strength at 60% and 80% overlap. For clinical translation, we recommend removal and reposition of the screw when the overlap is in the range of 20% to 40% or less. In vitro specimen studies are needed to verify these preliminary findings.
More Related Videos
12:25C-arm-Free Simultaneous OLIF51 and Percutaneous Pedicle Screw Fixation in a Single Lateral Position
Published on: September 16, 2022
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Upward Impending Motion
To better comprehend how a screw jack functions, consider the completely unraveled thread as a block in contact with the...
Moment of a Force: Problem Solving
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...