Related Experiment Video
Updated: Aug 15, 2025

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
A novel method to calculate compliance and airway resistance in ventilated patients
1Professor Emeritus Medicine, Anesthesiology and Engineering, The George Washington University, 700 New Hampshire Ave, NW, Suite 510, Washington, DC, 20037, USA. gutier@gwu.edu.
Background:
The respiratory system's static compliance (Crs) and airway resistance (Rrs) are measured during an end-inspiratory hold on volume-controlled ventilation (static method). A numerical algorithm is presented to calculate Crs and Rrs during volume-controlled ventilation on a breath-by-breath basis not requiring an end-inspiratory hold (dynamic method).
Methods:
The dynamic method combines a numerical solution of the equation of motion of the respiratory system with frequency analysis of airway signals. The method was validated experimentally with a one-liter test lung using 300 mL and 400 mL tidal volumes. It also was validated clinically using airway signals sampled at 32.25 Hz stored in a historical database as 131.1-s-long epochs. There were 15 patients in the database having epochs on volume-controlled ventilation with breaths displaying end-inspiratory holds. This allowed for the reliable calculation of paired Crs and Rrs values using both static and dynamic methods. Epoch mean values for Crs and Rrs were assessed by both methods and compared in aggregate form and individually for each patient in the study with Pearson's R2 and Bland-Altman analysis. Figures are shown as median[IQR].
Results:
Experimental method differences in 880 simulated breaths were 0.3[0.2,0.4] mL·cmH2O-1 for Crs and 0[- 0.2,0.2] cmH2O·s· L-1 for Rrs. Clinical testing included 78,371 breaths found in 3174 epochs meeting criteria with 24[21,30] breaths per epoch. For the aggregate data, Pearson's R2 were 0.99 and 0.94 for Crs and Rrs, respectively. Bias ± 95% limits of agreement (LOA) were 0.2 ± 1.6 mL·cmH2O-1 for Crs and - 0.2 ± 1.5 cmH2O·s· L-1 for Rrs. Bias ± LOA median values for individual patients were 0.6[- 0.2, 1.4] ± 0.9[0.8, 1.2] mL·cmH2O-1 for Crs and - 0.1[- 0.3, 0.2] ± 0.8[0.5, 1.2] cmH2O·s· L-1 for Rrs.
Discussion:
Experimental and clinical testing produced equivalent paired measurements of Crs and Rrs by the dynamic and static methods under the conditions tested.
Conclusions:
These findings support to the possibility of using the dynamic method in continuously monitoring respiratory system mechanics in patients on ventilatory support with volume-controlled ventilation.
More Related Videos
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Respiratory Volumes and Capacities I
Cardiopulmonary Resuscitation II: ACLS Airway Management