Headache after transcatheter closure of atrial septal defect: An attempt to explain its origin in the pediatric

Sebastian Smerdziński1,2, Michał Gałeczka3, Filip Tyc3,4

  • 1Department of Congenital Heart Defects and Pediatric Cardiology, Silesian Center for Heart Diseases, Zabrze, Poland. smerdzinski@gmail.com.

Kardiologia Polska
|January 3, 2023
PubMed

Insights

Transient headache episodes (THE) after transcatheter atrial septal defect (ASD) closure are linked to balloon calibration and lack of nickel protection. These findings help identify patients at higher risk for post-procedural headaches.

Area of Science:

  • Cardiology
  • Interventional Cardiology
  • Pediatric Cardiology

Background:

  • Transcatheter closure is the preferred treatment for atrial septal defects (ASD).
  • Transient headache episodes (THE) affect approximately 5% of patients post-procedure, with unclear causes.

Purpose of the Study:

  • To identify risk factors for transient headache episodes (THE) in pediatric patients following transcatheter ASD closure.

Main Methods:

  • Retrospective analysis of 840 pediatric patients undergoing transcatheter ASD closure with nitinol devices.
  • Logistic regression analysis to assess risk factors including age, weight, ASD diameter, device size, nitinol coating, fluoroscopy time, balloon calibration, device oversizing, and residual shunt.

Main Results:

  • 4.8% of patients experienced THE.
  • Higher incidence of THE associated with balloon calibration (82.5% vs. 43.3%), larger device/balloon sizes, and increased residual shunt.
  • Predictors of THE included age, balloon calibration, lack of nickel protection, fluoroscopy duration, and device oversizing.

Conclusions:

  • Balloon calibration during percutaneous ASD closure is a significant risk factor for early post-procedural headaches.
  • The absence of a protective layer against nickel release on nitinol devices is also linked to headache occurrence.