The Effect of Cycloplegia on the Biometer for Optical Low-coherence Reflectometry
1Ophthalmology, Medicana International Samsun Hospital, Samsun, Turkey.
Purpose:
The aim of this study was to compare the biometric measurements and different formulas for calculating intraocular lens (IOL) power by using the Lenstar LS900 biometer for optical low-coherence reflectometry before and after induction of cycloplegia in the adult population.
Materials And Methods:
In this cross-sectional study, 168 eyes of 168 healthy volunteers aged 40 - 86 years (59.22 ± 11.57) were included. Biometric measurements, including axial length (AL), anterior chamber depth (ACD), keratometry (K1 and K2), and white-to-white (WTW) were compared using a Lenstar LS900 optical biometer before and after induction of cycloplegia with 1% cyclopentolate. The IOL power was also compared using six different formulas (Barrett Universal II, Haigis, SRK/T, Hoffer Q, Holladay, and SRK-II) for the AcrySof MA60AC IOL before and after induction of cycloplegia.
Results:
There were no statistically significant differences in AL, K1 and K2, or WTW measurements before and after induction of cycloplegia. There was a significant increase only in ACD from the biometric parameters after the induction of cycloplegia (p < 0.05). Despite this change, there were no significant changes in IOL power calculations using the six different formulas before and after induction of cycloplegia.
Conclusions:
This study demonstrated that IOL power measurements using the Lenstar LS900 can be performed after cycloplegia.