Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients
Marie-Anne Blanchet1, Gabriel Mercier1, Antoine Delobel1
1Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada.
Background:
An accurate SpO value is critical in order to optimally titrate oxygen delivery to patients and to follow oxygenation guidelines. Limited prospective data exist on real-world performance of pulse oximeters in critically ill patients. The objective of this study was to assess accuracy and bias of the SpO values measured by several oximeters in hospitalized subjects.
Methods:
We included stable adults in the ICU with an arterial catheter in place. Main exclusion criteria were poor SpO signal and SpO > 96%. In each subject, we simultaneously evaluated 4 oximeters: Nonin (Plymouth, Minnesota) embedded in the FreeO2 device (OxyNov, Québec City, Québec, Canada), Masimo (Radical-7, Masimo, Irvine, California), Philips (FAST, Philips, Amsterdam, the Netherlands), and Nellcor (N-600, Medtronic, Minneapolis, Minnesota). Arterial blood gases were drawn and simultaneously each oximeters' SpO values were collected. SpO values were compared to the reference (arterial oxygen saturation [SaO ] value) to determine bias and accuracy. The ability for oximeters to detect hypoxemia and the impact of oximeters on oxygen titration were evaluated.
Results:
We included 193 subjects (153 male, mean age 66 y) in whom 211 sets of measurements were performed. The skin pigmentation evaluated by Fitzpatrick scale showed 96.2% of subjects were light skin (types 1 and 2). One oximeter overestimated SaO (Philips, +0.9%), whereas the 3 others underestimated SaO (Nonin -3.1%, Nellcor -0.3%, Masimo -0.2%). SaO was underestimated with Nonin oximeter in 91.3% of the cases, whereas it was overestimated in 55.2% of the cases with Philips oximeter. Moderate hypoxemia (SaO 86-90% or PaO 55-60 mm Hg) was detected in 92, 33, 42, and 11% of the cases with Nonin, Nellcor, Masimo, and Philips, respectively.
Conclusions:
We found significant bias and moderate accuracy between the tested oximeters and the arterial blood gases in the studied population. These discrepancies may have important clinical impact on the detection of hypoxemia and management of oxygen therapy.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
07:51Standardized Hemorrhagic Shock Induction Guided by Cerebral Oximetry and Extended Hemodynamic Monitoring in Pigs
Published on: May 21, 2019
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring pulse
Pulse
The pulse serves as a clinical...