Related Experiment Video
Updated: Aug 15, 2025

Delivery of In Vivo Acute Intermittent Hypoxia in Neonatal Rodents to Prime Subventricular Zone-derived Neural Progenitor Cell Cultures
Published on: November 2, 2015
Comparison of two automated oxygen controllers in oxygen targeting in preterm infants during admission: an
Hylke H Salverda1,2, Janneke Dekker3, Enrico Lopriore3
1Willem-Alexander Children's Hospital, Department of Pediatrics, Division of Neonatology, Leiden University Medical Center, The Netherlands, Leiden University Medical Center, Leiden, The Netherlands H.H.Salverda@lumc.nl.
Objective:
To compare the effect of two different automated oxygen control devices on time preterm infants spent in different oxygen saturation (SpO2) ranges during their entire stay in the neonatal intensive care unit (NICU).
Design:
Retrospective cohort study of prospectively collected data.
Setting:
Tertiary level neonatal unit in the Netherlands.
Patients:
Preterm infants (OxyGenie 75 infants, CLiO2 111 infants) born at 24-29 weeks' gestation receiving at least 72 hours of respiratory support between October 2015 and November 2020.
Interventions:
Inspired oxygen concentration was titrated by the OxyGenie controller (SLE6000 ventilator) between February 2019 and November 2020 and the CLiO2 controller (AVEA ventilator) between October 2015 and December 2018 as standard of care.
Main Outcome Measures:
Time spent within SpO2 target range (TR, 91-95% for either epoch) and other SpO2 ranges.
Results:
Time spent within the SpO2 TR when receiving supplemental oxygen was higher during OxyGenie control (median 71.5 [IQR 64.6-77.0]% vs 51.3 [47.3-58.5]%, p<0.001). Infants under OxyGenie control spent less time in hypoxic and hyperoxic ranges (SpO2<80%: 0.7 [0.4-1.4]% vs 1.2 [0.7-2.3]%, p<0.001; SpO2>98%: 1.0 [0.5-2.4]% vs 4.0 [2.0-7.9]%, p<0.001). Both groups received a similar FiO2 (29.5 [28.0-33.2]% vs 29.6 [27.7-32.1]%, p=not significant).
Conclusions:
Oxygen saturation targeting was significantly different in the OxyGenie epoch in preterm infants, with less time in hypoxic and hyperoxic SpO2 ranges during their stay in the NICU.
More Related Videos
07:52Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...