Radiation exposure in cardiac computed tomography imaging in Mie prefecture in 2021
Suguru Araki1, Kakuya Kitagawa2, Takanori Kokawa1
1Department of Radiology, Mie University Hospital, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Purpose:
Several effective radiation dose reduction methods have been developed for coronary computed tomography angiography (CTA); however, their use in daily clinical practice remains unknown. We aimed to investigate radiation exposure and the utilization of dose-saving strategies for coronary CTA in hospitals in Mie Prefecture, Japan.
Materials And Methods:
Image acquisition details and dose reports of 30 consecutive cardiac CT examinations performed in 2021 were obtained from 18 hospitals. The inclusion criteria were patients aged 20-80 years who weighed 50-70 kg and underwent coronary CTA using ≥ 64-row multidetector CT. The doses for the overall cardiac CT examination and coronary CTA were analyzed using the dose-length product (DLP) and CT dose index (CTDIvol), respectively. Multivariate analysis was performed to determine independent predictors that affect the radiation dose in coronary CTA.
Results:
The median DLP of cardiac CT was 774 (interquartile range [IQR]: 538-1119) mGy*cm, and the median CTDIvol of coronary CTA was 33 (IQR: 25-48) mGy. The 75th percentile values of DLP for cardiac CT and that of CTDIvol for coronary CTA were slightly lower than the values recorded in the Japan Diagnostic Reference Level (DRLs) 2020 report (1285 mGy*cm and 66.4 mGy, respectively) but were substantially higher than those reported in a previous large international dose survey (402 mGy*cm and 24 mGy, respectively). Iterative reconstruction was performed during all examinations. Only six hospitals (33%) used a low tube potential (≤ 100 kVp), and nine hospitals (50%) used electrocardiogram-triggered prospective scanning. Multivariate analysis revealed low heart rate, low tube potential, and use of electrocardiogram-triggered prospective scanning as independent predictors of CTDIvol ≤ 24 mGy (p < 0.001, respectively).
Conclusion:
As of 2021, low tube potential and prospective scanning are underutilized, whereas iterative reconstruction is used in every coronary CTA in Mie Prefecture. Further efforts to optimize the radiation exposure from cardiac CT scans are necessary.
More Related Videos
07:08Establishment of a Robust and Reproducible Model of Radiation-Induced Skin and Muscle Fibrosis
Published on: August 31, 2022
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...