Related Experiment Video
Updated: Aug 15, 2025

Continuous-wave Thulium Laser for Heating Cultured Cells to Investigate Cellular Thermal Effects
Published on: June 30, 2017
Study of Temperature Changes around Fibres in Super Pulse Thulium Fibre Laser in Vitro Lithotripsy
Ding Tianfu1, Xiao Bo1, Zeng Xue1
1Department of Urology, Tsinghua Changgung Hospital, Tsinghua University, Affiliated with the School of Clinical Medicine, Tsinghua University, Beijing 102218.
Purpose:
To investigate temperature changes around the fibres of the super pulse thulium fibre laser (SP-TFL) during in vitro lithotripsy.
Materials And Methods:
Stones were placed in the in vitro model. The laser was continuously excited for 180 s; the probe was positioned 5 mm around the fibre; the laser power was set at 10, 15, 20, 25, and 30 W; and the irrigation rate was set at 0, 15, 25, 35 ml/min, using a domestic SP-TFL. The temperature variations around the fibre under different power settings and different irrigation rates were compared.
Results:
Without irrigation, the temperature around the fibre rapidly reached the safety threshold of 43℃ at 24 s. At irrigation rate 15 ml/min and laser power <15 W, the temperature around the fibre was <43℃. Once the laser power increased to ≥20 W, the temperature around the fibre increased to >43℃ at its lowest plateau. At irrigation rate 25 ml/min and laser power ≤25 W, the temperature around the fibre was <43℃. At irrigation rate 35 ml/min and laser power <30 W, the fibre temperature was <43℃. When laser power was ≥30 W, the fibre temperature was >43℃.
Conclusion:
In extracorporeal ureteroscope SP-TFL lithotripsy, when the laser power is ≤15 W, ≤25 W, and ≤30 W, the irrigation rate should be maintained at ≥15 ml/min, ≥25 ml/min, and ≥35 ml/min, respectively.