A Novel Discretization Procedure in the CSI-FEM Algorithm for Brain Stroke Microwave Imaging
Valeria Mariano1, Jorge A Tobon Vasquez1, Francesca Vipiana1
1Department of Electronics and Telecommunications, Politecnico di Torino, 10129 Torino, Italy.
Sensors (Basel, Switzerland)
|January 8, 2023
Summary
A new discretization method improves microwave imaging accuracy for brain stroke detection. This contrast source inversion technique enhances focus on affected tissues, simplifying implementation and boosting precision in medical diagnostics.
Area of Science:
- Medical imaging
- Electromagnetics
- Computational methods
Background:
- Microwave imaging offers non-invasive diagnostic potential for conditions like brain stroke.
- Accurate reconstruction of dielectric properties is crucial for effective diagnosis.
- Current contrast source inversion (CSI) methods face challenges in implementation complexity and accuracy.
Purpose of the Study:
- To develop a novel contrast source variable discretization for improved microwave imaging.
- To enhance the accuracy and simplify the implementation of CSI algorithms.
- To apply and evaluate the proposed method for brain stroke detection using realistic 3D models.
Main Methods:
- Combining the contrast source inversion (CSI) method with the finite element method (FEM).
- Developing and implementing a novel contrast source variable discretization technique.
- Creating a synthetic environment with modeled antennas and coaxial ports for simulations.
- Reconstructing dielectric properties of human head tissues for brain stroke imaging.
Main Results:
- The novel discretization simplifies algorithm implementation and improves accuracy of discretized quantities.
- The proposed method demonstrates better focus on the stroke region compared to the standard CSI implementation in a realistic 3D scenario.
- Performance evaluation shows comparable average results but improved localization with the new discretization within a limited permittivity range.
Conclusions:
- The novel contrast source variable discretization offers enhanced accuracy and implementation simplicity for microwave imaging.
- The method shows promise for improving the resolution and focus in brain stroke detection.