Related Experiment Video
Updated: Aug 15, 2025

An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Multi-Element UWB Probe Optimization for Medical Microwave Imaging
Youness Akazzim1,2, Otman El Mrabet2, Jordi Romeu1
1School of Telecommunication Engineering, Universitat Politècnica de Catalunya, 08034 Barcelona, Spain.
Abstract:
The need for non-ionizing techniques for medical imaging applications has led to the use of microwave signals. Several systems have been introduced in recent years based on increasing the number of antennas and frequency bandwidth to obtain high resolution and good accuracy in locating objects. A novel microwave imaging system that reduces the number of required antennas for precise target location appropriate for medical applications is presented. The proposed system consists of four UWB extended gap ridge horn (EGRH) antennas covering the frequency band from 0.5 GHz to 1.5 GHz mounted on a cylindrical phantom that mimics the brain in an orthogonal set of two EGRH probes. This configuration has the ability to control both the longitudinal and transversal dimensions of the reconstructed target's image, rather than controlling the spatial resolution, by increasing the frequency band that can be easily affected by medium losses. The system is tested numerically and experimentally by the detection of a cylindrical target within a human brain model.