REDUCTION OF THE COMPUTED TOMOGRAPHY DOSE TO THE GONADS OF ADOLESCENT AND YOUNG ADULT MALES THROUGH USE OF A
Hiroyuki Kosaka1,2, Mikoto Tamura1, Shota Hattori1,2
1Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, 377-2 Ohno-higashi, Osakasayama, Osaka 589-8511, Japan.
Abstract:
This study aims to measure the shielding effect of a novel tungsten rubber sheet (TRS) on the exposure of male gonads to ionizing radiation during upper abdominal and abdominal plus pelvic computed tomography (CT) examinations. The air kerma at the gonad with and without shielding was measured using an anthropomorphic body phantom. Gonads were shielded using: (1) 360° wrap with TRS (0.5-mm thick) and (2) 180° wrap with TRS. The air kerma at a position of male gonads in a pelvic scan was 21.8 mGy, even when the gonads were in the off-axis range. The TRS reduced the dose to the male gonads in abdominal plus pelvic CT examinations by 61% and 38% for the 360° and 180° TRS wrap, respectively.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...