Modification by an aspiration catheter for vessel stretching in thrombectomy using a stent retriever in vitro
Yasuhiko Nariai1, Tomoji Takigawa1, Akio Hyodo1
1Department of Neurosurgery, Dokkyo Medical University Saitama Medical Center, 2-1-50 Minamikoshigaya, Koshigaya-shi, Saitama 343-8555, Japan.
Objectives:
Post-thrombectomy subarachnoid hemorrhage, which is caused by vessel stretching when retrieving a stent retriever (SR), is a complication of endovascular mechanical thrombectomy (EVT) using an SR. This in vitro study examined the association between EVT for middle cerebral artery (MCA) occlusion and M2 vessel stretching depending on the positioning of the SR or combined use of SR and aspiration catheter (AC) technique (CBT).
Materials And Methods:
A silicone vascular model was used. The maximum migration distance (MMD) in the anteroposterior view of an X-ray impermeable metal marker located at an M2 trunk 20 mm distal to the M1-2 bifurcation in angiographic images during an SR retrieval was calculated. According to the positioning of SR deployment (distal or proximal) and use of an AC [SR alone without an AC (SR alone) group, stent retrieval into an AC (SA) group, and stent-AC retrieval as a unit into a guiding catheter (SA unit) group], 60 attempts were conducted with each group of 10 attempts.
Results:
In distal SR deployment, the MMD in the CBT group, consisting of SA and SA unit groups, was significantly smaller than the SR alone group (1.36 ± 0.32 mm VS 2.39 ± 1.10 mm; p = 0.001). In proximal SR deployment, the CBT group showed a tendency to decrease MMD (1.38 ± 0.33 mm VS 1.63 ± 0.28 mm; p = 0.077).
Conclusions:
This in vitro study showed that CBT might help modify the stretching of M2 when an SR is distally deployed in MCA occlusion.