Intra-Operative Neural Monitoring of Thyroid Surgery in a Porcine Model
Published on: February 11, 2019
NIM TriVantage™ electromyogram endotracheal tube kinking and the preclusion using a tube stand in patients undergoing
Kensuke Suzuki1, Shingo Kawashima1,2, Atsushi Kobayashi1,2
1Department of Anesthesiology and Intensive Care, Hamamatsu University School of Medicine, Hamamatsu, Japan.
Background:
The NIM TriVantage™ electromyogram endotracheal tube (NIM [Nerve Integrity Monitoring] tube) is positioned at the midline of the mouth, as the electrodes must attach to both sides of the vocal cords for neural monitoring during thyroid surgery. The tube possibly kinks below the slip joint, resulting in ventilatory disturbance. We examined whether kinking of the NIM tube altered ventilation in a simulation model and assessed whether using a tube stand during thyroid surgery prevented such kinking and associated ventilatory instability.
Methods:
We assessed the impact of bending warmed (32°C) NIM tubes (7.0- and 8.0-mm ID, n = 10 per group) on airway pressure and measured tidal volume within a simulator. Additionally, we investigated whether a custom-designed tube stand could mitigate ventilatory impairment when employing a midline-fixed NIM tube in patients undergoing thyroid surgery.
Results:
In a simulator, 75° bending of the NIM (7.0 mm) tube significantly augmented airway pressure (31 [28 to 34] cmH2O, median [IQR], P < 0.001), and 60° and 75° bending of it caused a slight, but significant decrease in the measured tidal volume compared with 0° bending. Using our custom tube stand prevented kinking of the 7.0-mm NIM tube, with no clinically meaningful changes in airway pressure or delivered tidal volume during surgery.
Conclusions:
The NIM (7.0 mm) tube is vulnerable to kinking-induced ventilatory impairment in the clinical setting, and a tube stand prevents this issue. We recommend using a tube stand during thyroid surgery to ensure safe and effective neural monitoring.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques