Background:
Lead failure is the major limitation in implantable cardioverter-defibrillator (ICD) therapy. Long-term follow-up data for Biotronik Linox ICD leads are limited. Therefore, we analyzed the performance of all these leads implanted at our institution.
Materials And Methods:
All Linox and Linox Smart ICD leads implanted between 2006 and 2015 were identified. Lead failure was defined as electrical dysfunction (oversensing, abnormal impedance, exit block). Lead survival was described, according to Kaplan-Meier. Associations between lead failure and specific variables were examined. p < .05 was considered significant.
Results:
We included 417 ICD leads. The median follow-up time for Linox (n = 205) was 81 months and for Linox Smart (n = 212) 75 months. During that follow-up time, 30 Linox (14.6%) and 16 Linox Smart leads (7.6%) showed a malfunction. The 5-year lead survival probability was 97.4% for Linox and 95.2% for Linox Smart (log-rank test, p = .19). The 6- and 8-year lead survival probability for Linox was 93.6% and 84.6%, and for Linox Smart 93% and 91.9%. The only factor significantly associated with lead failure was younger patient age at implantation (hazard ratio/year: 0.97, 95% CI: 0.94-0.99, p = .002).
Conclusion:
This relatively large study with a long follow-up period highlights a relevant failure rate of Biotronik Linox leads. The performance of Linox versus Linox Smart ICD leads was comparable. Although we show an acceptable 5-year lead survival probability, we observed a marked drop after just 1 more year of follow-up. In an era of improving heart failure survival probability, a prolonged follow-up of ICD leads is increasingly clinically relevant.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:05Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Cardiac Catheterization III: Left Heart Catheterization
Cardiopulmonary Resuscitation III: AED Use