Related Experiment Video
Updated: Aug 14, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Benchmarking of stray neutron fields produced by synchrocyclotrons and synchrotrons in compact protontherapy centers
Gonzalo F García-Fernandez1, Eduardo Gallego1, José M Gómez-Ros2
1Departamento de Ingeniería Energética, ETSI Industriales, Universidad Politécnica de Madrid, José Gutiérrez Abascal 2, 28006, Madrid, Spain.
Abstract:
Proton therapy is an external radiotherapy using proton beams with energies between 70 and 230 MeV to treat some type of tumours with outstanding benefits, due to its energy transfer plot. There is a growing demand of facilities taking up small spaces and Compact Proton Therapy Centers (CPTC), with one or two treatment rooms, supposing the technical response of manufacturers to this request. A large amount of stray radiation is yielded in the interaction of proton beam used in therapy, neutrons mainly, hence, optimal design of shielding and verifications must be carried out in commissioning phases. Currently, almost 50 proton centers are under construction and start up in several countries, including ten in Spain. In the present work the effectiveness of shielding in two CPTC was verified with the Monte Carlo code MCNP6 by calculating the ambient dose equivalent, H*(10) due to secondary neutrons, outside the enclosures and walls of the center. The facilities modelled were the two centers currently operating in Spain, the first, since December 2019, with a superconductor synchrocyclotron, and the second, since March 2020, with a compact synchrotron. The geometry and materials are based on dimensions proposed a priori by the vendors, therefore, the paper is focused on check the suitability of the materials and thickness of the walls of the centers. Several models of the radiation sources were simulated, starting from a conservative assumptions, followed by more realistic scenarios. In all cases, the results reached for the ambient dose equivalent, H*(10), were below 1 mSv/year, which is the legal limit considered for the public in international references. Finally, considering that the recent ICRU Report 95 proposes changes in the operational quantities, the dose outside shieldingt has been evaluated in terms of the new next area surveillance quantity, H*, known as ambient dose, in the process of implementation.
More Related Videos
10:10Neutron Crystallography Data Collection and Processing for Modelling Hydrogen Atoms in Protein Structures
Published on: December 1, 2020
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Related Concept Videos
Nuclear Stability
To hold positively charged protons together...
Nuclear Binding Energy
Atomic Nuclei: Nuclear Spin State Population Distribution
Nuclear Transmutation