[Determination of platinum anticancer drugs in occupational settings by ICP-MS]
Abstract:
Objective: To establish an inductively coupled plasma mass spectrometry (ICP-MS) for platinum antineoplastic drugs in the environment. Methods: The platinum antineoplastic drugs in the environmental table were eluted by wiping and collecting pure water, and the supernatant was taken by centrifugation and inductively coupled plasma mass spectrometry for detection. Results: The concentration range of 0-8.0 μg/L was good, the correlation coefficient was 1.000, the detection limit was 0.0006 μg/L, the lower quantitative limit was 0.002 μg/L, the method precision was between 0.9%-1.3%, and the sample standard recovery rate was between 97.0%-98.5%. Conclusion: This method has low detection limit, high accuracy and precision, and simple sample pretreatment, which is suitable for the determination of platinum antineoplastic drugs in environmental tables.
More Related Videos
07:20Amide Coupling Reaction for the Synthesis of Bispyridine-based Ligands and Their Complexation to Platinum as Dinuclear Anticancer Agents
Published on: May 28, 2014
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab