Is TKA femoral implant stability improved by pressure applied cement? a comparison of 2 cementing techniques
Mareike Schonhoff1, Nicholas A Beckmann2, Martin Schwarze1
1Laboratory of Biomechanics and Implant Research, Department of Orthopaedics, Heidelberg University Hospital, 69118, Heidelberg, Germany.
Background:
The majority of knee endoprostheses are cemented. In an earlier study the effects of different cementing techniques on cement penetration were evaluated using a Sawbone model. In this study we used a human cadaver model to study the effect of different cementing techniques on relative motion between the implant and the femoral shaft component under dynamic loading.
Methods:
Two different cementing techniques were tested in a group of 15 pairs of human fresh frozen legs. In one group a conventional cementation technique was used and, in another group, cementation was done using a pressurizing technique. Under dynamic loading that simulated real life conditions relative motion at the bone-implant interface were studied at 20 degrees and 50 degrees flexion.
Results:
In both scenarios, the relative motion anterior was significantly increased by pressure application. Distally, it was the same with higher loads. No significant difference could be measured posteriorly at 20°. At 50° flexion, however, pressurization reduced the posterior relative motion significantly at each load level.
Conclusion:
The use of the pressurizer does not improve the overall fixation compared to an adequate manual cement application. The change depends on the loading, flexion angle and varies in its proportion in between the interface zones.