Thoracic motion-compensated cone-beam computed tomography in under 20 seconds on a fast-rotating linac: A simulation
Samuel J Blake1, Owen Dillon1, Hilary L Byrne1
1ACRF Image X Institute, Sydney School of Health Sciences, Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia.
Background:
Rapid kV cone-beam computed tomography (CBCT) scans are achievable in under 20 s on select linear accelerator systems to generate volumetric images in three dimensions (3D). Daily pre-treatment four-dimensional CBCT (4DCBCT) is recommended in image-guided lung radiotherapy to mitigate the detrimental effects of respiratory motion on treatment quality.
Purpose:
To demonstrate the potential for thoracic 4DCBCT reconstruction using projection data that was simulated using a clinical rapid 3DCBCT acquisition protocol.
Methods:
We simulated conventional (1320 projections over 4 min) and rapid (491 projections over 16.6 s) CBCT acquisitions using 4D computed tomography (CT) volumes of 14 lung cancer patients. Conventional acquisition data were reconstructed using the 4D Feldkamp-Davis-Kress (FDK) algorithm. Rapid acquisition data were reconstructed using 3DFDK, 4DFDK, and Motion-Compensated FDK (MCFDK). Image quality was evaluated using Contrast-to-Noise Ratio (CNR), Tissue Interface Width (TIW), Root-Mean-Square Error (RMSE), and Structural SIMilarity (SSIM).
Results:
The conventional acquisition 4DFDK reconstructions had median phase averaged CNR, TIW, RMSE, and SSIM of 2.96, 8.02 mm, 83.5, and 0.54, respectively. The rapid acquisition 3DFDK reconstructions had median CNR, TIW, RMSE, and SSIM of 2.99, 13.6 mm, 112, and 0.44 respectively. The rapid acquisition MCFDK reconstructions had median phase averaged CNR, TIW, RMSE, and SSIM of 2.98, 10.2 mm, 103, and 0.46, respectively. Rapid acquisition 4DFDK reconstruction quality was insufficient for any practical use due to sparse angular projection sampling.
Conclusions:
Results suggest that 4D motion-compensated reconstruction of rapid acquisition thoracic CBCT data are feasible with image quality approaching conventional acquisition CBCT data reconstructed using standard 4DFDK.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography