Comparison of five different methodologies for evaluating ankle-foot orthosis stiffness
Benjamin R Shuman1,2, Deema Totah3, Deanna H Gates4
1Center for Limb Loss and Mobility, VA Puget Sound, 1660 S Columbian Way, Seattle, WA, USA. bshuman@hjf.org.
Journal of Neuroengineering and Rehabilitation
|January 22, 2023
Summary
Standardizing mechanical testing for ankle-foot orthoses (AFOs) is crucial. Different testing methods yield significantly varied stiffness measurements, impacting clinical use and research for AFOs.
Area of Science:
- Biomechanics
- Orthotics and Prosthetics Engineering
Background:
- Ankle-foot orthosis (AFO) mechanical properties significantly influence user gait mechanics.
- Current methodologies for evaluating AFO mechanical properties lack standardization.
Purpose of the Study:
- To compare the outcomes of five distinct evaluation frameworks for assessing AFO stiffness.
- To highlight the impact of non-standardized testing on AFO mechanical property measurements.
Main Methods:
- Five different testing methods were employed to evaluate 13 carbon composite AFOs.
- Four methods used custom test fixtures for dorsiflexion in the sagittal plane; one involved quasi-static deflection via hanging weights.
- Rotational stiffness was calculated, and differences between methods were statistically analyzed.
Main Results:
- Significant differences in measured AFO stiffness were observed across the five methods.
- Methods with less constrained motion (e.g., hanging weights) yielded lower stiffness values.
- The relative difference in stiffness measurements between methods ranged from 13% to 156%.
Conclusions:
- The study underscores the critical need for standardized mechanical testing protocols for AFOs.
- Methodological variations lead to substantial discrepancies in stiffness measurements, potentially limiting clinical application.
- Standardization is essential for reliable AFO stiffness metrics in clinical prescription and research.