Vitamin D: marker, measurand & measurement
Niek F Dirks1,2, Etienne Cavalier3, Annemieke C Heijboer4,5,6,7
1Atalmedial Diagnostics Centre, Spaarne Gasthuis, Haarlem, The Netherlands.
Accurate vitamin D status assessment requires understanding vitamin D metabolite measurements. Automated assays may misinterpret results for some patients, especially those using vitamin D2, making liquid chromatography-tandem mass spectrometry a more reliable option.
Area of Science:
- Clinical Chemistry
- Endocrinology
- Nutritional Science
Background:
- Vitamin D metabolites are crucial for assessing vitamin D status and diagnosing calcium homeostasis disorders.
- Laboratories commonly measure total 25-hydroxyvitamin D (25(OH)D), with some offering separate measurements of 25(OH)D2 and 25(OH)D3, and other metabolites like 1,25-dihydroxyvitamin D and 24,25-dihydroxyvitamin D.
Purpose of the Study:
- To provide an updated overview of key vitamin D metabolism markers.
- To define intended measurands for vitamin D assays.
- To discuss the pros and cons of automated assays versus liquid chromatography-tandem mass spectrometry (LC-MS/MS).
Main Methods:
- Review of current literature on vitamin D metabolite assays.
- Comparative analysis of automated immunoassays and LC-MS/MS techniques.
- Discussion of assay limitations and interpretation pitfalls.
Main Results:
- Automated assays may not accurately measure 25(OH)D in all patient populations, particularly those supplemented with vitamin D2.
- LC-MS/MS offers a more reliable measurement of 25(OH)D, especially when properly developed and standardized.
- Understanding assay-specific limitations is vital for accurate interpretation of vitamin D status.
Conclusions:
- Automated assays have limitations in accurately quantifying 25(OH)D for certain patient groups, including those using vitamin D2.
- LC-MS/MS provides a more dependable method for measuring 25(OH)D when validated and standardized.
- Clinical laboratories must be aware of assay pitfalls to ensure accurate patient diagnosis and management of vitamin D status.
More Related Videos
10:07Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact
Published on: February 10, 2015
08:01Measurement of Insulin- and Contraction-Stimulated Glucose Uptake in Isolated and Incubated Mature Skeletal Muscle from Mice
Published on: May 16, 2021
Related Concept Videos
01:24Tools of Measurement
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Measurement: Derived Units
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Interval Level of Measurement
Data measured using the interval scale are similar to ordinal level data because they have a definite arrangement. However, in the interval level of measurement, the differences between data values are meaningful even though the data does not have a starting point.
Temperature is measured using the interval scale. It is measurable data, and the difference between...
Uncertainty in Measurement: Reading Instruments