Related Experiment Video
Updated: Aug 13, 2025

Robotic D3 Partial Duodenal Resection with Primary Side-to-Side Anastomosis
Published on: December 15, 2023
A comparison between the da Vinci Xi EndoWrist Stapler and a conventional laparoscopic stapler in rectal transection:
Nan Zun Teo1, James Chi-Yong Ngu1
1Department of General Surgery, Changi General Hospital, Singapore, Singapore.
Background:
Distal rectal transection following robotic total mesorectal excision for rectal cancer is challenging. This can be performed with either a robotic stapler (RS) or laparoscopic stapler (LS). We compared the operative outcomes and ergonomic differences between RS and LS.
Methods:
Forty patients with rectal cancer who were planned for robotic rectal surgery were randomized to either RS or LS.
Results:
The mean number of stapler cartridges expended was similar between RS and LS (1.75 vs. 2.05, p = 0.082). Significantly less stapler adjustments were required in the RS group (1.55 vs. 2.75, p = 0.014). Results from the user experience survey showed that RS scored better.
Conclusion:
The RS required less adjustments and resulted in a better user experience. While this complemented the use of robotics in rectal surgery, it did not result in a significant reduction in the number of stapler cartridges required for rectal transection.