Related Experiment Video
Updated: Aug 12, 2025

A Multi-detection Assay for Malaria Transmitting Mosquitoes
Published on: February 28, 2015
Handheld Purification-Free Nucleic Acid Testing Device for Point-of-Need Detection of Malaria from Whole Blood
Aneesh Kshirsagar1, Gihoon Choi1, Vishaka Santosh2
1School of Electrical Engineering and Computer Science, Pennsylvania State University, University Park, Pennsylvania16802, United States.
Abstract:
World Health Organization's aim to eliminate malaria from developing/resource-limited economies requires easy access to low-cost, highly sensitive, and specific screening. We present a handheld nucleic acid testing device with on-chip automated sample preparation to detect malaria (Plasmodium falciparum) infection from a whole blood sample as a feasibility study. We used a simple two-reagent-based purification-free protocol to prepare the whole blood sample on a piezo pump pressure-driven microfluidic cartridge. The cartridge includes a unique mixing chamber for sample preparation and metering structures to dispense a predetermined volume of the sample lysate mixture into four chambers containing a reaction mix. The parasite genomic DNA concentration can be estimated by monitoring the fluorescence generated from the loop-mediated isothermal amplification reaction in real time. We achieved a sensitivity of ∼0.42 parasite/μL of whole blood, sufficient for detecting asymptomatic malaria parasite carriers.