Retinal Vascular Reactivity as Assessed by Optical Coherence Tomography Angiography
Published on: March 26, 2020
Adaptive dynamic analysis-based optical coherence tomography angiography for blood vessel projection artifact
Junxiong Zhou1,2, Yuntao Li1,3,2, Jianbo Tang1
1Guangdong Provincial Key Laboratory of Advanced Biomaterials, Department of Biomedical Engineering, Southern University of Science and Technology, Shenzhen, Guangdong 518055, China.
Abstract:
Optical coherence tomography angiography (OCTA) for blood vessel 3-D structure imaging suffers from blood vessel projection artifacts/tail artifacts when using a long decorrelation time (e.g., repeat B-scan acquisition in regular OCTA) or loss of micro vessel signal when using a short decorrelation time. In this work, we developed an adaptive first-order field autocorrelation function (g1) analysis-based technique to suppress the projection artifacts under macro vessels while enhancing the dynamic signal of micro vessels. The proposed method is based on the differences of the decorrelation rate and the phase variations of g1 between the vessel voxels and the artifacts regions. A short or long decorrelation time was applied to obtain the dynamic index of the projection artifacts region or the blood vessel region, respectively. Compared to the slab subtraction-based post-image processing-based techniques, the proposed approach addresses this problem on a physical basis and shows the ability to suppress the projection artifacts while enhancing the detection of the micro vessels.