Evaluation of implantable collamer lens sizing developed by reviewing the horizontal compression-vault coefficient
Tomoaki Nakamura1, Tomoya Nishida, Naoki Isogai
1From the Nagoya Eye Clinic, Nagoya, Japan (Nakamura, Nishida, Isogai, Sugiyama, Yoshida); Department of Ophthalmology, Keio University School of Medicine, Tokyo, Japan (Kojima).
Purpose:
To evaluate the implantable collamer lens (ICL)-sizing method using the partial regression coefficient of the implanted ICL size to review the conventional horizontal compression coefficient and match the results of clinical observation.
Setting:
Nagoya Eye Clinic, Nagoya, Japan.
Design:
Interventional case series.
Methods:
Patients who underwent ICL V4c implantation to correct myopia and myopic astigmatism were enrolled. The stepwise multiple regression analysis used achieved vault as a dependent variable and preoperative biometric parameters as explanatory variables. The partial regression coefficient of the implanted ICL size was obtained to develop the optimal ICL-sizing formula, the Nakamura-Kojima formula version 3 (NK-formula V3). 85 eyes of 45 patients were implanted with an ICL size recommended by the NK-formula V3. At 3 months postoperatively, the achieved vault was measured using anterior segment optical coherence tomography to validate the NK-formula V3.
Results:
The study enrolled 174 patients (174 eyes). The partial regression coefficient of the implanted ICL size adopted as the compression-vault coefficient in the multiple regression equation predicting the vault was 0.729. To validate the NK-formula V3, 77 (90.6%), 7 (8.2%), and 1 (1.2%) eyes were in the moderate-vault, low-vault, and high-vault categories, respectively. The mean difference between the achieved and predicted vaults was 0.064 ± 0.190 (range: -0.264 to 0.742) mm for the NK-formula V3 and 0.176 ± 0.217 (range: -0.254 to 0.907) mm for the NK-formula V2.
Conclusions:
As in vivo coefficient measurement experiments are not possible, the partial regression coefficient is the best option for developing a formula to predict the optimal ICL size.