Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
597
Systematic Review of Fitbit Charge 2 Validation Studies for Exercise Tracking.
1Nell Hodgson Woodruff School of Nursing, Emory University, Atlanta, GA.
Translational Journal of the American College of Sports Medicine
|January 30, 2023
Summary
The Fitbit Charge 2 (FBC2) shows accurate heart rate (HR) monitoring at low to moderate exercise intensities. However, its step count accuracy is inconclusive, limiting its suitability for all populations in research and clinical settings.
Area of Science:
- Wearable technology validation
- Biomedical engineering
- Health informatics
Background:
- Research-grade devices like Polar chest straps (ECG) and ACTiGraph accelerometers are validated for heart rate (HR) and step count, respectively.
- Wearable activity trackers measuring HR and steps concurrently have shown conflicting validation results against research-grade equipment.
Purpose of the Study:
- To review validation studies of the Fitbit Charge 2 (FBC2) for accuracy in measuring HR and step count.
- To evaluate the FBC2's reliability for use by researchers and clinicians.
Main Methods:
- A systematic review was conducted following PRISMA guidelines.
- Eight eligible articles published between 2018-2019 were identified from multiple databases and hand-searches.
- Concordance correlation coefficients (CCC) and mean absolute percent error (MAPE) were used to assess agreement and error between the FBC2 and criterion devices.
Main Results:
- FBC2 HR monitoring agreement (CCC) ranged from 26% to 92%, decreasing with increased exercise intensity.
- Step count agreement (CCC) increased from 38% to 99% with higher exercise intensity.
- HR error (MAPE) ranged from 9.21% to 68%, increasing with exercise intensity.
- Step count error (MAPE) was 12% in healthy adults but 46% in older adults with heart failure.
Conclusions:
- The FBC2 demonstrates acceptable HR accuracy at low-to-moderate exercise intensities.
- Step count validation remains inconclusive, requiring further research with controlled testing.
- The FBC2 may be suitable for HR monitoring in specific populations but is not definitively validated for step counting in all users.